当前位置::首页 > 技术支持 > 电压击穿试验仪-安装条件 技术文章

电压击穿试验仪-安装条件

点击次数:1929  更新时间:2017-06-22

50KV电压击穿试验仪 安装条件:

 电源: 220V±10%的单相交流电压和50Hz±1%的频率

电源电压稳定度外界电源电压波动≤10%

长×宽×高700mm×800mm×1650mm

设备自重:    100kg

运行环境温度:15 ~ 30℃,相对湿度:30%~85%能够稳定运行。

接地要求仪器需要单独接地,接地附合国家标准要求

接地电阻要求≤4Ω(用户实验室自行准备)

 

 

电压击穿试验仪组成:
1、升压部件:由调压器和高压变压器组成0~50KV的升压部分。
2、动部件:由步进电机均匀调节调压器使加给高压变压器的电压变化。
3、检测部件:由集成电路组成的测量电路。通过信号线把检测的模拟信号和开关信号传给计算机。
4、计算机软件:通过智能电路把由检测设备采集的测控信号传给计算机。计算机根据采集的信息控制设备运行并处理试验结果。
5、试验电极:根据国家标准(1408.1-2006)随设备提供三个电极,具体规格为:Ф25mm×25mm两个;Ф75mm×25mm一个。(订做除外)

 

 

 

标签:

电压击穿强度试验仪

耐电压强度击穿试验仪

电气强度绝缘试验机

介电击穿强度试验仪

 

 

 

50 kv voltage puncture tester Installation conditions:

Power: 220 v + / - 10% of the single-phase ac voltage and the frequency of 50 hz + 1%

Power supply voltage stability the outside power supply voltage fluctuation 10% or less

Length x width x height of 700 mm * 800 mm * 1650 mm

Equipment weight: 100 kg

Running environment temperature: 15 ~ 30 ℃, relative humidity: 30% ~ 85% can stable operation.

Grounding requires instrument need separate grounding, along with the national standard requirements

Grounding resistance 4 or less Ω (user laboratory prepare)

 

 

The breakdown voltage of tester:
1, booster parts: consists of a voltage regulator and the high voltage transformer 0 ~ 50 kv booster section.
2, moving parts: uniform regulation voltage regulator by stepper motor to high voltage transformer voltage change.
3, test parts: measurement circuit composed of integrated circuit.Through the analog signal and the detection line switch signal to the computer.
4, computer software, by testing equipment acquisition through inligent circuit of measurement and control signals to the computer.Computer control equipment operation according to the collected information and process the test results.
5, test electrode: according to the national standard (1408.1-2006) with the equipment to provide three electrodes, specifications are: Ф two 25 mm x 25 mm;Ф a 75 mm x 25 mm.Except (build)

联系人:陈丹
手机:
18911395947